PR325A Thermocouple Calibration Furnace has excellent performance and rich functions. It adopts a new structural design, has a longer service life, and solves the problems of furnace positioning and high temperature electric leakage through the built-in metal positioner.
PR325A Thermocouple Calibration Furnace has excellent performance and rich functions. It adopts a new structural design, has a longer service life, and solves the problems of furnace positioning and high temperature electric leakage through the built-in metal positioner.
The control part uses part of the technology of the PR330 Multi-zone Temperature Calibration Furnace, which has the ability to adjust the axial temperature uniformity slightly. Compared with the traditional thermocouple calibration furnace, better verification or calibration results can be obtained without a isothermal block.
I. Features
No need for a isothermal block, and the axial temperature uniformity over the full range is better than 1°C/6cm
The controller can automatically adjust the balance power at both ends, and can obtain 1°C/6cm axial temperature uniformity without a isothermal block in the temperature range of 300°C~1200°C, which can effectively reduce the uncertainty of the verification or calibration process.
Integrated high-accuracy temperature controller and reference compensator
Using PR2601 temperature controller, it has a measurement accuracy of 0.01. With the special reference end compensator, the accuracy is better than 0.6℃+0.1%RD when using type N temperature-controlled thermocouple.
Built-in positioner for easy sensor positioning
The bottom of the built-in metal positioner is 32cm away from the test end of the furnace mouth, and the furnace loading operation can be completed by simply inserting the sensor into the bottom of the positioner.
High temperature electric leakage suppression
The ground terminal is reserved externally, and after connecting the metal positioner, the influence of leakage at high temperature on the electrical measuring instrument can be effectively suppressed.
Longer service life
Under the same operating conditions, by increasing the load capacity of the internal heating wire, the service life of several times the traditional calibration furnace can be obtained.
Rich software and hardware functions
Using the front color touch screen, it can display and set general measurement and control parameters, and can also perform operations such as timed power on and off, temperature stability settings, and WIFI settings.
II. Other Functions
Other Functions | |
Temperature control sensor multi-temperature point correction Adaptive temperature control parameters Real-time temperature, power curve display Built-in reference junction compensation | Custom temperature fluctuation calculation Custom alarm temperature upper and lower limits Bluetooth, WIFI expandable Optional units °C, °F, K |
III. Technical Parameters
Product selection and technical parameters
Model | PR325A | Remarks |
Working temperature range | 300℃~1200℃ | / |
Furnace cavity dimension | φ40mm×600mm | / |
Temperature control accuracy | 0.5℃, when ≤500℃ 0.1%RD, when>500℃ | Geometric center point temperature of the furnace cavity |
60mm axial temperature field uniformity | ≤1.0℃ | 300℃~1200℃ Furnace cavity geometric center ±30mm |
Radial temperature field uniformity | ≤0.4℃ | Furnace cavity geometric center |
Temperature stability | ≤0.3℃/10min | / |
General technical parameters
Item | Parameters |
Dimensions | 700×370×500mm (L×W×H) |
Display screen | 4.0-inch industrial touch screen with a resolution of 800×480 pixels |
Communication method | RS232 (Standard), WiFi, Bluetooth (Optional) |
Weight | 55kg |
Rated power | 3kW |
Power supply | 220VAC±10% |
Working environment | -5~35℃, 0~80%RH, Non-condensing |
Storage environment | -20~70℃, 0~80%RH, Non-condensing |
Changsha Panran Technology Co., Ltd.